Sensorex
DO1200 Galvanic Dissolved Oxygen Sensor
DO1200 Galvanic Dissolved Oxygen Sensor
0 in stock
Description
The DO1200 galvanic dissolved oxygen sensors measure the oxygen content of water using an electrochemical measurement principle. PTFE membrane cap included.
Built in an economical but durable 12mm epoxy body, these sensors are suitable for portable or benchtop use. The DO1200 can also be installed inline for continuous monitoring applications.
Specifications
Reads | Dissolved Oxygen |
Range | 0 − 20 mg/L |
Accuracy | +/– 0.05 mg/L |
Response time | ~0.3 mg/L/per sec |
Temperature range °C | 1 − 50 °C |
Max pressure | 7,5 bar (100PSI) |
Max depth | 70m (1,157 ft) |
Connector | Male BNC |
Cable length | 1 meter |
Internal temperature sensor | No |
Time before recalibration | ~1 Year |
Life expectancy | ~4 Years |
Maintenance | ~18 Months |
- Galvanic (electrochemical) measurement principle
- Replaceable membrane cap and fill solution
- 12mm durable epoxy body
- For benchtop, portable, or inline use
Shipping & Returns
It may happen that you want to return an order. Because the product does not suit you, maybe there is another reason why you would not want the order. Whatever the reason, you have the right to cancel your order up to 14 days after receipt without giving a reason.
After cancellation you have 14 days to return your product. You will then be credited 85% order amount excluding shipping. For more information click the following link: /pages/shipping-and-returns
What's Included?
The DO1200 Galvanic Dissolved Oxygen Sensor comes with:
• 1x 1x PTFE membrane cap